Skip to main content
. 2012 Nov 7;12:157. doi: 10.1186/1471-230X-12-157

Table 1.

The differences of esophageal motility measurements between the 22-channel WPM and 36-sensor SSM during wet swallows (n = 12)

  22-channel WPM 36-sensor SSM Z value P
LES proximal margin (cm)
41.5(38-43)
42(39-43)
-2.11
0.035
LES length (cm)
4(4-4)
3.5(3-4)
0.00
1.000
LES rest pressure (mmHg)
13.5(10-21)
16.5(14-22.5)
-1.69
0.091
LES relax ratio (%)
97.5(92-98.5)
84(77-88.5)
-2.20
0.028
Distal esophageal body contraction intensity (mmHg)
68.5(40-85.25)
80(50-126.5)
-1.88
0.060
Distal esophageal body contraction duration (s)
3.85(2.9-4.2)
3.7(3.4-4.525)
-1.18
0.238
Proximal esophageal body contraction intensity (mmHg)
39(22-54)
51.5(40-62.75)
-1.96
0.050
Proximal esophageal body contraction duration (s)
3.05(2.6-3.275)
3.2(2.5-3.325)
-0.67
0.505
UES proximal margin (cm)
17(16-18)
16.5(16-17.25)
-0.81
0.417
UES length (cm)
4(4-5)
4(3-5)
-0.52
0.603
UES rest pressure (mmHg)
56.5(43-82.5)
77.5(60-113)
-1.88
0.060
UES relax ratio (%) 100(100-100) 100(97-100) -0.76 0.446