Skip to main content
. 2012 Nov 21;7(11):e49516. doi: 10.1371/journal.pone.0049516

Table 2. System suitability-Sensitive Assay: Excellent accuracy and linearity.

BoNT/A Dilutional linearity//Recovery Table for Sensitive Assay
Concentration Expected Relative Potency Actual Relative Potency Confidence Interval (CI) Percent Recovery Number of assays Number of assay with CI overlapping 1
0.57× 1.75 1.69 1.54–1.85 96.3 4 0 of 4
0.66× 1.5 1.50 1.36–1.66 100 4 0 of 4
0.8× 1.25 1.34 1.26–1.41 106.8 4 0 of 4
1.33× 0.75 0.85 0.78–0.91 111.9 4 0 of 4
1.5× 0.66 0.68 0.63–0.73 103.3 4 0 of 4
2.0× 0.5 0.55 0.52–0.59 111.6 4 0 of 4