Skip to main content
. 2012 Nov 27;7(11):e50299. doi: 10.1371/journal.pone.0050299

Table 1. Comparison of cellular uptake and electrochemical byproducts in the three compartments of the exposure device.

Compartment Uptakea ROSb (DCFintensity) pHc
Anode 3.14±1.53 4.94±2.22 5.5
Center 1.07±0.08 0.71±0.32 7.5
Cathode 1.09±0.12 0.76±0.33 12
a

Uptake of Dextran-FITC in terms of exposed/constitutive uptake.

b

ROS determined by DCF relative fluorescence in terms of exposed/unexposed cells.

c

Measured in HBSS after its exposure to LEF.