Table 2. Evaluation of rapid test kits for POCT at each clinic.
Clinic | Sensitivity | Specificity | PPVa | NPVb | ||||
SEA-IC | PDM-IC | SEA-IC | PDM-IC | SEA-IC | PDM-IC | SEA-IC | PDM-IC | |
A (n = 15) | 90.9 | 83.3 | 100 | 100 | 100 | 100 | 80.0 | 90.0 |
B (n = 10) | 50.0 | 60.0 | 100 | 100 | 100 | 100 | 57.1 | 71.4 |
C (n = 56) | 100 | 100 | 89.3 | 94.7 | 90.3 | 90.0 | 100 | 100 |
D (n = 60) | 82.2 | 90.5 | 100 | 94.4 | 100 | 97.4 | 65.2 | 81.0 |
E (n = 11) | 100 | 75.0 | 80.0 | 100 | 85.7 | 100 | 100 | 87.5 |
F (n = 16) | 100 | 100 | 100 | 85.7 | 100 | 90.0 | 100 | 100 |
G (n = 21) | 88.2 | 93.3 | 25.0 | 66.7 | 83.3 | 87.5 | 33.3 | 80.0 |
H (n = 28) | 90.9 | 87.5 | 76.5 | 100 | 71.4 | 100 | 92.9 | 95.2 |
I (n = 16) | 100 | 100 | NAc) | 100 | 75.0 | 100 | NA | 100 |
J (n = 15) | 75.0 | 60.0 | 85.7 | 100 | 85.7 | 100 | 75.0 | 83.3 |
K (n = 16) | 100 | 100 | 66.7 | 100 | 92.9 | 100 | 100 | 100 |
L (n = 32) | 95.5 | 100 | 60.0 | 92.3 | 84.0 | 95.0 | 85.7 | 100 |
M (n = 36) | 71.4 | 91.7 | 81.8 | 100 | 71.4 | 100 | 81.8 | 96.0 |
Total (n = 332) | 89.2 | 92.5 | 80.5 | 95.3 | 87.9 | 94.9 | 82.4 | 93.1 |
Correlationcoefficient | 0.706 | 0.657 | 0.180 | 0.784 | ||||
P value | 0.007 | 0.020 | 0.555 | 0.003 |
Positive predictive value.
Negative predictive value.
Not applicable.