Skip to main content
. 2011 Feb;40(2):115–122. doi: 10.1259/dmfr/31708191

Table 4. Equivalent dose (µSv) to tissues and organs in the head and neck from Kodak 9500 large field of view (FOV) (18 × 21 cm) and medium FOV (9 × 15 cm) examinations using default exposure settings for small, medium and large adults and comparing pre-production and current production units.

Remainder
Field (Adult size) Unit status Bone marrow Thyroid Oesophagus Skin Bone surface Salivary glands Brainb Braina Lymphatic nodesb Extrathoracic airwaya Musclea,b Oral mucosab
18 × 21 cm (small) Added filtration 134 471 36 70 485 1513 1017 1017 63 1262 63 1437
18 × 21 cm (medium) Pre production 292 2000 216 345 1181 4573 2030 2030 190 3736 190 4239
Added filtration 218 835 131 123 747 2645 1640 1640 114 2286 114 2549
% reduction 25% 58% 39% 65% 37% 42% 19% 19% 40% 39% 40% 40%
18 × 21 cm (large) Pre production 399 2045 307 451 1548 5447 2730 2730 233 4669 233 5219
Added filtration 342 1356 233 182 1098 4161 2694 2694 185 3655 185 4048
% reduction 14% 34% 24% 60% 29% 24% 1% 1% 21% 22% 21% 22%
9 × 15 cm (small) Pre production 150 960 89 132 1039 3672 180 180 155 2738 155 3572
Added filtration 70 475 52 33 255 1582 100 100 68 1196 68 1530
% reduction 53% 51% 42% 75% 75% 57% 44% 44% 56% 56% 56% 57%
9 × 15 cm (medium) Pre production 173 1010 117 176 1202 4682 265 265 189 3386 189 4413
Added filtration 113 533 54 55 386 1680 264 264 94 1738 94 2143
% reduction 35% 47% 54% 68% 68% 64% 0% 0% 50% 49% 50% 51%
9 × 15 cm (large) Added filtration 170 840 90 86 547 3612 350 350 156 2836 156 3570

aICRP 2007; bICRP 1990