Table 3. Effective organ dose and total effective dose (µSv) for the different scanning protocols of the NewTom 9000 (Quantitative Radiology, Verona, Italy).

Scan number Bone marrow Thyroid Oesophagus Skin Bone surface Salivary glands Brain Remainder tissues/organs
Total
Lymphatic nodes Extrathoracic region Muscles Oral mucosa
1 9.4 31.0 2.4 0.6 1.9 15.5 1.8 0.8 13.6 0.8 17.5 95.3
2 7.9 33.9 2.8 0.5 1.6 14.0 1.6 0.7 12.3 0.7 15.8 91.8
3 7.8 29.3 2.4 0.4 1.5 13.3 1.2 0.7 11.8 0.7 15.4 84.5
4 10.9 15.9* 1.4* 0.6 2.2 15.9 2.2 0.8 13.7 0.8 17.6 82.0
5 10.0 16.6* 1.4* 0.5 2.0 14.8 1.6 0.8 13.2 0.8 17.4 79.1

Shielding methods: 1, without collar around the neck; 2, with one collar loosely on the front of the neck; 3, with two collars loosely on the front and back of the neck; 4, with one collar tightly on the front of the neck; 5, with two collars tightly on the front and back of the neck.