Published in final edited form as: Phys Med Biol. 2012 Oct 9;57(21):6981–6997. doi: 10.1088/0031-9155/57/21/6981

Table 4.

Results for the 3-dimensional γ-index for the IMRT test case in Fig. 4b. The comparison uses δd0 = 3 mm, δD0 = 3%, and region of interest denned by 30% of the maximum dose. IMRT uses collimators to shape the dose distribution which tests the limits of the numerical techniques. In this case, even in the presence of collimators, both the 1st-order Method F and 2nd-order Method I give Pγ≤1 with ~ 1% accuracy.

Pγ≤1 [%] Accuracy [%] Calculation time [s]
A 90.80 - 22000
B 89.88 1.0 2300
C 37.97 58 1.5
D 81.65 10.1 36.7
E 91.80 1.1 4.7
F 91.75 1.1 28.0
G 90.71 0.1 34.7
H 88.71 2.3 51.9
I 90.04 0.8 35.6