. 2012 Nov 26;39(12):7338–7349. doi: 10.1118/1.4765459

Table 4.

Dimensions and volumes of the in vivo ablation zones, estimated from t43≥ 240 min isodose boundary calculated by computer models (T restricted to ∼ 100 °C), after 10 and 15 min ablations with applicators in configurations 1–3. (NC = noncontiguous).

    mbl = 5 kg m−3 s−1
mbl = 12 kg m−3 s−1
mbl = 15 kg m−3 s−1
Time Configuration d1 (cm) d2 (cm) d3 (cm) V (cm3) d1 (cm) d2 (cm) d3 (cm) V (cm3) d1 (cm) d2 (cm) d3 (cm) V (cm3)
Three transducers (each 1.2 mm OD × 10 mm L) per catheter
10 min                          
  1 4.5 2.8 4.0 33.8 3.7 1.6 3.4 16.3 NC NC NC 6.1
  2 4.0 3.8 3.7 34.7 4.1 3.9 3.9 37.6 3.6 3.3 3.6 24.8
  3 5.6 4.8 4.0 62.2 5.6 4.6 4.1 57.3 NC NC 4.0 42.5
15 min                          
  1 5.0 3.5 4.4 47.3 4.2 2.5 3.8 27.4 3.7 1.5 3.4 15.1
  2 4.5 4.4 4.3 50.4 4.6 4.5 4.4 53.1 4.2 4.0 4.1 40.6
  3 6.1 5.6 4.5 87.6 6.0 5.4 4.4 80.7 5.7 4.9 4.1 65.3
Four transducers (each 1.2 mm OD × 10 mm L) per catheter
10 min                          
  1 4.5 2.9 5.2 46.7 4.2 2.3 5.0 35.6 3.8 1.6 4.8 23.8
  2 4.0 3.9 4.9 48.0 4.1 3.9 5.1 51.3 3.6 3.3 4.7 34.4
  3 5.7 4.9 5.2 85.8 5.6 4.6 5.3 79.0 NC NC 5.1 58.9
15 min                          
  1 5.0 3.6 5.6 65.2 4.6 3.1 5.3 51.2 4.3 2.6 5.1 39.2
  2 4.6 4.5 5.5 68.9 4.6 4.5 5.6 71.8 4.2 4.1 5.2 55.3
  3 6.2 5.7 5.5 118.4 6.1 5.5 5.5 110.0 5.8 5.0 5.4 90.0