. 2012 Oct 1;3(6):529–533. doi: 10.4161/viru.21540

Table 2. Problem handling.

  Step Problem Possible reason Solution
1
Reagents setup
We observed loss of PCR efficiency after multiple freezing-thawing cycles.
Primer degradation
To avoid degradation, aliquoted primer mixes should not be thawed more than 3–4 times. Small portions (50–100 μl) of primer mixes should be kept at -20°C.
2
Reaction setup
Uniformity
Reaction volume is small (5 μl). Preparation of single reaction that requires pipeting of volumes below 0.5 μl is very difficult.
We mix enough mastermix to fill whole 96 well plate and run multiple reactions at the same time.
3 Reaction setup Uniformity Pipetting errors caused by manual pipetting and small reaction volume. We usually use an electronic pipette with the “dispense” mode.