Skip to main content
. Author manuscript; available in PMC: 2013 Jul 1.
Published in final edited form as: IEEE Trans Neural Syst Rehabil Eng. 2012 Jul;20(4):422–431. doi: 10.1109/TNSRE.2012.2197865

TABLE I.

Performance metrics for CUSUM, sGLRT and uGLRT detectors and their disjunctive combination.

Detector Average Rats with seizure onset detected prior to the first MJ (%) Median time from PTZ injection to seizure onset detection (s)
Accuracy (%) Sensitivity (%) Specificity (%)
CUSUM 74.47 23.06 91.9 61.54 56
sGLRT 78.31 21.47 97.66 61.54 46
uGLRT 75.32 1.22 98.72 23.08 118
Disjunctive combination 76.14 33.73 89.7 76.92 18