Precision and accuracy of Letrozole in rat serum

Spiked concentr. (μg mL−1) Intra-day Inter-day

Mean detected concentr. (μg mL−1) RSD (%) RE (%) Mean detected concentr. (μg mL−1) RSD (%) RE (%)
0.5 0.45 ± 0.05 10.14 –10.36 0.44 ± 0.05 10.69 –11.52
20 19.22 ± 0.49 2.53 –3.89 19.22 ± 0.53 2.77 –3.89
50 49.07 ± 0.87 1.77 –1.89 48.87 ± 0.82 1.68 –2.26

Intra-day: n=5; Inter-day: n=3 days with 5 replicates per day; Values were expressed as mean ± SD.