Skip to main content
. Author manuscript; available in PMC: 2014 Jul 1.
Published in final edited form as: Magn Reson Med. 2012 Aug 8;70(1):259–268. doi: 10.1002/mrm.24432

Table 2.

Q measurements for elements of the developed dual-nuclei array.

Coil Environment Qunloaded Qloaded Coil Noise (%)
Sodium receive coil 1 330 105 17
Sodium receive coil 2 260 85 18
Sodium receive coil 3 190 70 21
Sodium receive coil 4 180 70 22
Sodium birdcage 1* 220 40 10
Sodium birdcage 1 130 50 22
Sodium birdcage 4 110 55 29
Proton coil 1 210 140 42
Proton coil 4 205 155 51

All Q measurements were made after minor capacitive adjustments (less than 10% for the sodium birdcage, and less than 5% for the sodium array and proton array) to compensate for shielding during integration of the three modules such that all sodium coils were tuned to 78.6 MHz and proton coils were tuned to 297.2 MHz.

*

indicates that measurements were made before detuning diodes and RF chokes were added to the birdcage.