Skip to main content
. Author manuscript; available in PMC: 2013 Dec 21.
Published in final edited form as: Phys Med Biol. 2012 Nov 22;57(24):8135–8152. doi: 10.1088/0031-9155/57/24/8135

Table 1.

Target localization accuracy and generated artifacts for FDPM-based system.

Number of phase-delays N (−) Benchtop Gantry-two projections Gantry-four projections
128 2.38 1.41 1.41 rms error (mm)

0 0 0 artifacts (%)
64 2.38 1.41 1.41 rms error (mm)

37 28 0 artifacts (%)
32 2.65 1.38 1.39 rms error (mm)

64 36 0 artifacts (%)