Skip to main content
. 2012 Aug 21;54(1):174–181. doi: 10.1093/jrr/rrs067

Table 1.

Comparison of point doses for plans and measurement (cGy)

Plan Solid slab maximum dose (Diff)a Central axis maximum dose Eyelid dose (Diff)
MC 196.1 47.3 197.7 ± 10
EBT2 197 (+ 0.5%) 44 201d (+ 1.7%)
Pinnacle_CT#b 196.7 (+ 0.3%) 46.9 159.6 ± 4.5 (– 19.3%)
Pinnacle_W ROIc 202.8 (+ 3.4%) 47.0 161.7 ± 6.8 (– 18.2%)

aDiff = percentage of difference relative to the MC results.

bPinnacle_CT# = Pinnacle plan based on the modified CT images.

cPinnacle_W ROI = Pinnacle plan based on the contouring and density override for the tungsten shield and steel knob.

dThis eyelid dose is the maximum dose taken from EBT2.