Table 3.
Implant bending moment and axial force derived experimentally from TFAS (n = 6) and UCR (n = 6) implants and theoretically from FE-TFAS model
| Loading mode | Implant side | Implant bending moments (Nm) | Implant axial force (N) | ||||
|---|---|---|---|---|---|---|---|
| TFAS | UCR | FE-TFAS | TFAS | UCR | FE-TFAS | ||
| Flexion | Left | ||||||
| Mean | 2.40* | 0.72 | 0.21 | 99.8 | 72.3 | 25.1 | |
| SD | 1.46 | 0.25 | – | 66.1 | 35.8 | – | |
| Right | |||||||
| Mean | 1.42* | 0.45* | 0.38 | 59.2 | 57.3 | 50.8 | |
| SD | 0.59 | 0.24 | – | 40.6 | 39.8 | – | |
| Extension | Left | ||||||
| Mean | 1.32* | 2.70* | 0.85 | −24.1* | −145.0* | 11.6 | |
| SD | 1.05 | 0.69 | – | −23.4 | −71.8 | – | |
| Right | |||||||
| Mean | 0.87* | 2.58* | 0.79 | −38.6 | −168.3 | 3.2 | |
| SD | 0.53 | 0.61 | – | −33.78 | −61.1 | – | |
| Left axial rotation | Left | ||||||
| Mean | 1.48 | 1.12 | 0.22 | 23.9* | −24.8* | −22.5 | |
| SD | 0.99 | 0.81 | – | 49.0 | −55.6 | – | |
| Right | |||||||
| Mean | 0.80 | 0.90 | 1.44 | 47.56* | −42.3* | 157.9 | |
| SD | 0.51 | 0.64 | – | 56.3 | −57.7 | – | |
| Right axial rotation | Left | ||||||
| Mean | 1.80 | 1.52 | 2.04 | 73.1* | −67.1* | 187.4 | |
| SD | 1.15 | 0.51 | – | 44.4 | −36.7 | – | |
| Right | |||||||
| Mean | 0.85 | 1.13 | 0.40 | −8.9 | −37.1 | −45.2 | |
| SD | 0.59 | 0.50 | – | −37.7 | −53.7 | – | |
| Left lateral bending | Left | ||||||
| Mean | 1.43* | 2.88* | 0.50 | 9.5* | −141.6* | 16.8 | |
| SD | 1.07 | 0.94 | – | 41.8 | −66.4 | – | |
| Right | |||||||
| Mean | 1.00 | 0.72 | 1.15 | 42.6 | 36.7 | 76.6 | |
| SD | 0.45 | 0.49 | – | 39.3 | 56.9 | – | |
| Right lateral bending | Left | ||||||
| Mean | 1.87 | 0.78 | 1.01 | 75.3 | 68.0 | 82.1 | |
| SD | 1.34 | 0.6 | – | 68.2 | 59.2 | – | |
| Right | |||||||
| Mean | 0.67* | 2.20* | 0.28 | −6.3* | −108.7* | −3.2 | |
| SD | 0.43 | 0.48 | – | −33.4 | −44.3 | – | |
TFAS Total Facet Arthroplasty System, UCR posterior fusion system, FE finite element, SD standard deviation
* Significant difference between TFAS and UCR