Table 3.

Implant bending moment and axial force derived experimentally from TFAS (n = 6) and UCR (n = 6) implants and theoretically from FE-TFAS model

Loading mode Implant side Implant bending moments (Nm) Implant axial force (N)
TFAS UCR FE-TFAS TFAS UCR FE-TFAS
Flexion Left
 Mean 2.40* 0.72 0.21 99.8 72.3 25.1
 SD 1.46 0.25 66.1 35.8
Right
 Mean 1.42* 0.45* 0.38 59.2 57.3 50.8
 SD 0.59 0.24 40.6 39.8
Extension Left
 Mean 1.32* 2.70* 0.85 −24.1* −145.0* 11.6
 SD 1.05 0.69 −23.4 −71.8
Right
 Mean 0.87* 2.58* 0.79 −38.6 −168.3 3.2
 SD 0.53 0.61 −33.78 −61.1
Left axial rotation Left
 Mean 1.48 1.12 0.22 23.9* −24.8* −22.5
 SD 0.99 0.81 49.0 −55.6
Right
 Mean 0.80 0.90 1.44 47.56* −42.3* 157.9
 SD 0.51 0.64 56.3 −57.7
Right axial rotation Left
 Mean 1.80 1.52 2.04 73.1* −67.1* 187.4
 SD 1.15 0.51 44.4 −36.7
Right
 Mean 0.85 1.13 0.40 −8.9 −37.1 −45.2
 SD 0.59 0.50 −37.7 −53.7
Left lateral bending Left
 Mean 1.43* 2.88* 0.50 9.5* −141.6* 16.8
 SD 1.07 0.94 41.8 −66.4
Right
 Mean 1.00 0.72 1.15 42.6 36.7 76.6
 SD 0.45 0.49 39.3 56.9
Right lateral bending Left
 Mean 1.87 0.78 1.01 75.3 68.0 82.1
 SD 1.34 0.6 68.2 59.2
Right
 Mean 0.67* 2.20* 0.28 −6.3* −108.7* −3.2
 SD 0.43 0.48 −33.4 −44.3

TFAS Total Facet Arthroplasty System, UCR posterior fusion system, FE finite element, SD standard deviation

* Significant difference between TFAS and UCR