Main performance parameters for the developed sensor S1.

Parameter Value Unit % of RL
Measuring Range 5 N NA
Linearity 2.7 mN 0.06%
Hysteresis 10 mN 0.21%
Cross talk 5.0 mN 0.1%
Resolution * 2.0 mN 0.04%
*

Resolution is mainly limited by noise; for characterization, a high precision low noise power supply has been employed which shows peak to peak noise Vpp ≤ 0.3 mV.