Table 2.

Treatment adequacy by comparing the delivered and planned minimum dose (Dmin) to the clinic target volume (CTV) and the minimal dose received by the highest 95% of CTV (D95 of CTV) based on the target motions tracked during the first 3, 5, 10, and all treatment fractions. The treatment is considered adequate when the delivered Dmin is at least 95% of the planned Dmin of the CTV. Bold indicates inadequate treatment.

Patient
Number
Dmin ratio (Dmin, deliver/Dmin, plan) D95 ratio (D95deliver/D95plan)
3 fx 5 fx 10 fx All fx 3 fx 5 fx 10 fx All fx
1 1.00 1.00 1.01 1.00 1.00 1.00 1.00 1.00
2 1.00 1.00 1.00 0.97 1.00 1.00 1.00 1.00
3 0.98 0.96 0.94 0.97 1.00 1.00 1.00 1.00
4 0.95 0.96 0.96 0.98 1.00 1.00 1.00 1.00
5 0.99 0.99 0.99 0.99 1.00 1.00 1.00 1.00
6 0.98 0.99 0.97 0.93 1.00 1.00 1.00 1.00
7 0.89 0.91 0.98 1.00 1.00 1.00 1.00 1.00
8 0.94 0.96 0.99 1.00 1.00 1.00 1.00 1.00
9 0.88 0.92 0.94 0.93 1.00 1.00 1.00 1.00
10 0.90 0.90 0.90 0.84 1.00 1.00 1.00 0.99
11 0.75 0.70 0.73 0.79 0.99 0.99 0.99 1.00
12 1.00 1.00 1.01 1.00 1.00 1.00 1.00 1.00
13 0.93 0.95 0.99 0.96 1.00 1.00 1.01 1.01
14 0.99 1.00 1.00 1.00 1.00 1.00 1.00 1.00
15 1.00 0.96 0.98 0.98 1.01 1.01 1.01 1.01
16 0.90 0.86 0.87 0.86 1.00 1.00 1.00 1.00