Table 1.

Characteristics of breast tomosynthesis systems in clinical use or under development. The specifications for these systems, especially the prototypes, could change in the future.

System GE Essential Hologic Selenia Dimensions IMS Giotto TOMO Philips MicroDose Planmed Nuance Excel DBT Siemens MAMMOMAT Inspiration
Detector type Full field—indirect Full field—direct (a-Se) Full field—direct (a-Se) Linear slit scan–spectral photon counting (Si) Full field—direct (a-Se) Full field—direct (a-Se)
Detector size (cm) 24 × 30 24 × 29 24 × 30 21 line detectors, each 24 cm long 24 × 30 24 × 30
Detector pixel size (μm) 100 70 (binned 2 × 2)1 85 50 (perpendicular to motion) 85 85
Detector motion Static Rotating Static Continuous slit scan Rotating during exposure2 Static
X-ray tube target Mo or Rh W W W W W
X-ray tube filtration 0.03 mm Mo or 0.025 mm Rh 0.7 mm Al 0.05 mm Rh or 0.05 mm Ag 0.5 mm Al 0.075 mm Ag or 0.06 mm Rh 0.05 mm Rh
X-ray tube motion Step-and-shoot Continuous Step-and-shoot Continuous Continuous Continuous
Angular range (deg) 25 153 40 11 30 504
Number of projections 9 15 135 21 15 25
Scan time (s) 7 3.7 12 3–10 20 254
Source to detector distance (cm) 66 70 68 66 65 65.5
Detector to center of rotation distance6 (cm) 4 0 2 −40 4.37 4.7
Air gap (cm) 2.2 2.5 2.2 0.4–2.4 2.38 1.7
Reconstruction method Iterative FBP Iterative with total variation regularization Iterative Iterative FBP
Development stage7 Prototype Commercial system Commercial system8 Prototype Prototype Commercial system8
References 16, 31 16 32, 33, 34, 35 26, 27, 36 37 16, 38, 39, 40
1

Reconstruction results in an in-plane pixel size of approximately 100 μm.

2

Detector and compressed breast are rotated with x-ray source, while each projection is being acquired (x-ray tube current is on). Detector and breast are returned to original position at end of each projection.

3

This system supports an angular range up to 30°, which requires a software change by the manufacturer.

4

These values for angular range and scan time are for the x-ray tube motion; the x-ray exposures actually cover ∼46° in 20 s.

5

In the current commercial system outside the U.S., this system can be set to use 50% of the total tube output for the 0° projection and the rest distributed evenly among the other 12 projections.

6

Positive values are towards the x-ray source.

7

For systems which perform both mammography and DBT, this information is specific for the DBT mode.

8

Currently not approved for clinical use in the U.S. by the FDA.