Table 2.

Modeling of the BrainLAB Imaging couch top (ICT) in the Philips Pinnacle3 treatment planning system using different combinations of carbon fibers (CF) and foam densities (F) for 10 cm × 10 cm field size

  Percentage deviation of Pinnacle3treatment planning system predicted dose to measured dose
Angle
Measured dose (cGy)
No Couch
CF0.6
CF0.5
CF0.46,
CF0.46
CF0.6
CF0.55
CF0.7
 
 
 
F0.02
F0.03
F0.05
F0.1
F0.1
F0.03
F0.1
180
89.3
1.31
-1.04
-0.26
-0.93
-1.82
-2.83
-0.71
-3.28
170
89.0
0.94
-0.97
-0.30
-0.97
-1.87
-2.77
-0.64
-3.33
160
87.9
1.72
-0.79
-0.10
-0.79
-1.70
-2.72
-0.44
-3.18
150
85.9
1.68
-0.30
0.40
-0.41
-1.46
-2.51
0.05
-2.86
140
82.4
2.67
-0.36
0.49
-0.24
-1.58
-2.79
0.12
-3.28
130
76.7
4.40
0.23
1.27
0.36
-1.21
-2.77
0.75
-3.42
120
67.5
7.12
1.05
2.53
1.19
-0.73
-2.81
1.79
-3.55
110
58.0
5.83
-1.06
0.49
-0.03
-1.23
-3.47
-0.20
-3.82
100
64.5
-0.96
-0.96
-0.96
-0.96
-0.96
-0.96
-0.96
-0.65
Sum
 
24.72
-4.20
3.56
-2.77
-12.56
-23.63
0.23
-27.36
t-test   0.0059 0.07 0.27 0.18 0.00007 0.000005 0.99 0.00002

Note that the “measured dose (cGy)” is the dose measured as described in phantom dose measurement through the imaging couch top.