Table 5.

Validation of BrainLAB imaging couch top (ICT) modeled in iPlan RT Dose treatment planning system (TPS); comparing measured dose with TPS computed dose

(a) 10 cm × 10 cm field size
  iPlan RT Dose predicted dose (cGy) Percentage deviation from measured dose
Angle
Measured dose (cGy)
No ICT
With ICT
No ICT
With ICT
180
89.3
92.45
89.34
3.49
0.01
170
89.0
92.16
89.00
3.59
0.04
160
87.9
91.24
87.92
3.81
0.03
150
85.9
89.34
85.92
4.06
0.07
140
82.4
86.63
82.44
5.14
0.05
130
76.7
82.01
77.59
6.89
1.13
120
67.5
74.27
67.32
10.04
-0.26
110
58.0
63.17
58.55
8.89
0.92
100
64.5
65.42
65.43
1.40
1.42
(b) 5 cm × 5 cm field size
180
81.2
84.09
80.48
3.55
0.90
170
80.8
83.76
80.14
3.61
0.87
160
79.7
82.74
79.03
3.88
0.78
150
77.5
80.67
76.76
4.09
0.96
140
74.0
77.79
73.26
5.13
0.99
130
68.3
72.91
67.74
6.72
0.85
120
59.2
64.71
58.56
9.27
1.12
110
50.0
53.7
49.78
7.31
0.52
100 56.0 55.74 55.74 -0.55 0.55

Note that the “measured dose (cGy)” is the dose measured as described in phantom dose measurement through the imaging couch top.