Table 2.

Dosimetric constraints utilized for tomotherapy planning for pre-operative helical tomotherapy

  Index Expected value Definition
Target volume
 
 
 
-Planning Target Volume (PTV)
V95
>95%
Volume receiving at least 95% of prescribed dose
D95
>95% (51.3 Gy)
Dose received by 95% of the volume
D98
>90% (48.6 Gy)
Dose received by 98% of the volume
D2
<107% (57.8 Gy)
Dose received by 2% of the volume corresponding to the “hot spots”
D_mean
54 Gy
Average dose (close as possible to the prescribed dose)
Organs at risk
 
 
 
-Gastrointestinal [9]
D200cc
<50 Gy
Dose received by 200cc of the intestinal volume
D2
<54 Gy
Dose received by 2% of the volume corresponding to the “hot spots”
V45
<33%
Volume receiving 45 Gy
V20
<50%
Volume receiving 20 Gy
-Contralateral kidney
D2
<12 Gy
Dose received by 2% of the volume corresponding to the “hot spots”
-Spinal canal D2 <45 Gy Dose received by 2% of the volume corresponding to the “hot spots”