Skip to main content
. 2013 Feb;51(2):618–620. doi: 10.1128/JCM.02565-12

Table 1.

Accuracy of study results for Premier CAMPY EIA compared to those for the reference method

Study EIA result No. of samples
% (range)
Culture positive Culture negative Sensitivity Specificity PPV NPV
Initial verification study Positive 10 1 100 (66–100) 98 (88–100) 91 (57–100) 100 (91–100)
Negative 0 49
Study following implementationa Positive 3 4 75 (22–98) 96.5 (92–99) 42.9 (12–80) 99 (94–100)
Negative 1 111
a

Discrepant results resolved by PCR analysis.