. 2013 Feb;51(2):452–458. doi: 10.1128/JCM.01982-12

Table 3.

LoDs of two-step pan-HPeV and EV assays with or without MS2 internal control

Assay and viral stock dilution (no. of CCID50s/ml) Without MS2 ICc
With MS2 IC
No. detected/total no. (%) Mean CT ± SD No. detected/total no. (%) Mean CT ± SD Mean MS2 IC CT
Pan-HPeV assaya
    102.6 6/6 (100) 32.9 ± 1.4 6/6 (100) 31.3 ± 0.8 31.3
    101.6 6/6 (100) 36.1 ± 1.2 6/6 (100) 36.6 ± 1.5 31.4
    100.6 6/6 (100) 39.1 ± 0.8 1/6 (17) NAd 31.7
    100.06 0/6 (0) NA 0/6 (0) NA 31.6
EV assayb
    101.2 6/6 (100) 29.9 ± 0.2 6/6 (100) 30.7 ± 0.3 31.6
    100.2 6/6 (100) 32.9 ± 0.4 6/6 (100) 33.6 ± 0.4 31.8
    100.02 6/6 (100) 36.3 ± 0.9 6/6 (100) 37.3 ± 0.6 32.0
    100.002 0/6 (0) NA 0/6 (0) NA 31.9
a

The HPeV3 strain was obtained from ZeptoMetrix with an initial viral titer of 1 × 108.6 CCID50s/ml. The virus culture was serially diluted with universal viral transport medium and tested at appropriate viral titers.

b

The echovirus 9 strain was obtained from ATCC with an initial viral titer of 1 × 107.2 CCID50s/ml. The virus culture was serially diluted with universal viral transport medium and tested at appropriate viral titers.

c

IC, internal control.

d

NA, not applicable.