Skip to main content
. 2013 Jan 24;8(1):e53233. doi: 10.1371/journal.pone.0053233

Table 4. Intra- and Inter-assay precision of the total BNP systems.

Added BNP concentration pmol/L Measured concentration pmol/L CV Bias
Mean S.D. % %
Intra-assay (n = 5) 2.0 2.3 0.2 7.0 15.0
25 25 2.1 8.4 1.0
100 99 7.1 7.2 −0.7
Inter-assay (n = 15) 2.0 2.1 0.2 9.5 5.0
25 24 1.7 2.9 −4.0
100 100 1.9 1.9 0.0