Pass rate versus IMRT technique, treatment planning system, linear accelerator manufacturer, and linac-TPS combination.

      Criterion failed
  Pass rate (%) Attempts Dose DTA Dose and DTA
IMRT technique
Dynamic MLC 88 296 26 5 5
IMAT 86 103 11 0 3
Segmental 76 634 109 15 25
Solid attenuator 43 7 4 0 0
TomoTherapy
93
99
6
1
0
Treatment planning system
Eclipse 88 387 30 8 7
Pinnacle3 75 425 84 8 13
TomoTherapy 93 99 6 1 0
XiO 76 137 19 4 10
Other
78
91
17
0
3
Linear accelerator manufacturer
Elekta 67 130 37 4 2
Siemens 70 135 32 3 6
TomoTherapy 93 99 6 1 0
Varian
85
775
81
13
25
Linac-TPS combination
Elekta-Pinnacle3 66 90 28 3 0
Siemens-Pinnacle3 67 76 21 0 4
TomoTherapy-HiArt 93 99 6 1 0
Varian-Eclipse 90 372 22 7 7
Varian-Pinnacle3 81 267 38 5 9
Varian-XiO 77 74 10 1 6