Skip to main content
. 2010 Oct-Dec;2(4):207–214.

Table 4.

Reactivity of the designed ELISA assay with different panels of serum samples

Samples Designed assay Commercial kit

Positive Negative Positive Negative
Positive panel (n=198) 196 2 198 0
Negative panel (n=250) 1 249 0 250
Unselected panel (n=100) 7 93 7 93

Unselected panel refers to serum samples collected from different referral hepatitis diagnostic laboratories without prior information about their hepatitis B status. Positive and negative panels were collected from the Iranian Blood Transfusion Organization based on reactivity with the commercial ELISA kit which was employed as a gold standard