Skip to main content
. 2012 Jun 17;202(1):77–86. doi: 10.1007/s00430-012-0250-5

Table 3.

Dilution of lysate samples derived from culture supernatants at which the assays positively detect (s/co ≥1.0) specific HIV subtype antigen or antibody

HIV subtype Elecsys® HIV combi PT ADVIA centaur® HIV Ag/Ab combo AxSYM® HIV Ag/Ab combo ARCHITECT® HIV Ag/Ab combo
HIV antigen
 A 1:64,000 1:64,000 1:64,000 1:128,000
 B 1:37,280 1:18,640 1:37,280 1:37,280
 C 1:26,720 1:26,720 1:26,720 1:53,440
 D 1:21,280 1:21,280 1:21,280 1:42,560
 E 1:31,968 1:7,992 1:31,968 1:63,936
 F 1:15,936 1:15,936 1:7,968 1:15,936
 G 1:52,896 1:26,448 1:52,896 1:105,792
 H 1:11,376 1:11,376 1:11,376 1:22,752
 O 1:256 Failed to detect diluted antigen 1:128 1:512
 HIV-2 1:64 Failed to detect diluted antigen 1:8 Failed to detect diluted antigen
HIV antibodya
 A 1:170,000 1:17,000 1:17,000 1:17,000
 A 1:5,000 1:5,000 1:5,000 1:500
 B 1:15,000 1:1,500 1:1,500 1:1,500
 B 1:120,000 1:120,000 1:120,000 1:120,000
 C 1:100,000 1:10,000 1:100,000 1:10,000
 C 1:200,000 1:200,000 1:200,000 1:200,000
 D 1:50,000 1:5,000 1:5,000 1:5,000
 D 1:28,500 1:28,500 1:28,500 1:2,850
 E 1:50,000 1:500 1:5,000 1:500
 E 1:50,000 1:5,000 1:5,000 1:500
 F 1:22,800 1:228 1:2,280 1:2,280
 F 1:100,000 1:100,000 1:100,000 1:100,000
 G 1:35,000 1:35,000 1:35,000 1:3,500
 G 1:140,000 1:140,000 1:14,000 1:14,000
 O 1:10,000 Failed to detect diluted antibody 1:10,000 1:1,000
 O 1:500,000 Failed to detect diluted antibody 1:5,000 1:5,000
 HIV-2 1:50,000 1:500 1:5,000 1:500
 HIV-2 1:60,000 1:600 1:6,000 1:6,000

aTwo different antibody samples were tested for each subtype