Essential parameters in OCT

En face TD-OCT Longitudinal TD-OCT SB-OCT SS-OCT
Depth resolution is determined by: Optical source bandwidth Optical source bandwidth Optical source bandwidth Tuning bandwidth
Axial scan range is: Unlimited Unlimited Limited by the spectrometer resolution Limited by the optical source linewidth
Dynamic focus is: Practical Possible but difficult at high speed A-scan line rates Nonapplicable Nonapplicable
Sensitivity versus OPD is: Constant Constant Maximum in OPD = 0 Maximum in OPD = 0
Mirror terms None None Manifest if OPD = 0 is crossed Manifest if OPD = 0 is crossed
Line rate determined by: Transversal scanner Axial scanner Reading the linear camera Tuning the swept source
Maximum achievable line rate 16 kHz (using a resonant scanner at 8 kHz) ∼100 kHz ∼300 kHz ∼5 MHz
Time to create an en face image of 500 lines each of 500 pixels along the line at the maximum achievable line rate is: Given by the time to produce a C-scan frame, 31 ms Given by the time to acquire a volume made from 500 B-scan frames of 500 line scans, ∼2.5 s Given by the time to acquire a volume made from 500 B-scan frames of 500 line scans, ∼0.83 s Given by the time to acquire a volume made from 500 B-scan frames of 500 line scans, ∼50 ms
Time to acquire a volume of 500 × 500 × 500 pixels at the maximum achievable line rate is: 15.5 s ∼2.5 s ∼0.83 s ∼50 ms