26 |
Tissue is leaky/dye is spreading |
Poor perfusion/perfusate |
Ensure that the perfusate does not contain sodium chloride, is not heated above 53° C, is of pH between 7.2 and 7.4 and is prepared the day of perfusion |
|
|
Too long time elapses between opening the chest cavity and start of perfusion |
Practice perfusions until there is minimal delay between opening the chest and initiating the perfusion, with no more than a few seconds elapse between impaling the needle in the left ventricle, initiating flow, and opening the right atrium. |
|
|
Postfixation too short for specific tissue |
Increase the postfixation time |
28 |
Dye enters the soma but doesn’t seem to spread into the distal dendrites |
Postfixation too long for specific tissue |
Decrease the postfixation time |
29 |
Dye doesn’t |
Tip is clogged (for LY, tip will appear red) |
With tip away from slice but inside PBS, apply opposite current (i.e. positive current) of 10–40 nA for a few seconds. If tip still clogged, switch to a new glass micropipette |
|
|
Micropipette tip too small |
Adjust puller to give pipettes with slightly less resistance |
|
|
Tip is resting on membrane or nucleus |
Tap manipulator to unclog tip; also try very minute readjustments of pipette |
29 |
Current injector is beeping and no dye is coming out |
The current circuit is broken |
Check entire current path; if platinum is coated, remove coat with a razor blade; if the platinum wire has kinks in it, replace it; for air objective setups, ensure that there is enough liquid to create a current path between micropipette and ground |
26–30 |
Dendrites appear beaded |
Too long time elapses between opening of chest cavity and start of perfusion |
See troubleshooting for step 26 |
|
|
A non-fixative (e.g. PB) was used before perfusate |
DO NOT use any non-fixative such as saline to “clear” the blood-vessels; this increases the duration of hypoxia of the brain and starts the apoptotic/necrotic process |
47 |
Can only visualize and image dendrites that are less than 30 μm below the surface |
Coverslip is slanted |
Remount |
|
|
Objective spring is broken |
Send in objective for repair |
49 |
Dendrite appears to be moving across the screen during confocal imaging |
Nail polish has not had enough time to harden or not enough nail polish has been applied |
Because the recommended mounting media is aqueous and therefore never hardens, the stability of the slide prep relies on sufficient and hardened nail polish; wait 2–3 days before imaging; if movement is still seen, apply another coat of nail polish and wait another couple of days |
49,58 |
Some dendrites or portions of dendrites appear fuzzy in the confocal image |
Air bubbles in the immersion oil |
Remove oil from slide and add a fresh amount of oil |
|
|
Air bubbles in mounting media |
Remount |
|
|
Dendrite is under a blood vessel |
Move on to the next segment; do not use for analysis |
56–61 |
Problems with NeuronStudio |
|
Consult the online manual http://research.mssm.edu/cnic/help/ns/index.html
|