Skip to main content
. Author manuscript; available in PMC: 2013 Feb 7.
Published in final edited form as: Nat Protoc. 2011 Aug 25;6(9):1391–1411. doi: 10.1038/nprot.2011.389

Table 2.

Troubleshooting table.

Step Problem Possible reason Solution
26 Tissue is leaky/dye is spreading Poor perfusion/perfusate Ensure that the perfusate does not contain sodium chloride, is not heated above 53° C, is of pH between 7.2 and 7.4 and is prepared the day of perfusion
Too long time elapses between opening the chest cavity and start of perfusion Practice perfusions until there is minimal delay between opening the chest and initiating the perfusion, with no more than a few seconds elapse between impaling the needle in the left ventricle, initiating flow, and opening the right atrium.
Postfixation too short for specific tissue Increase the postfixation time
28 Dye enters the soma but doesn’t seem to spread into the distal dendrites Postfixation too long for specific tissue Decrease the postfixation time
29 Dye doesn’t Tip is clogged (for LY, tip will appear red) With tip away from slice but inside PBS, apply opposite current (i.e. positive current) of 10–40 nA for a few seconds. If tip still clogged, switch to a new glass micropipette
Micropipette tip too small Adjust puller to give pipettes with slightly less resistance
Tip is resting on membrane or nucleus Tap manipulator to unclog tip; also try very minute readjustments of pipette
29 Current injector is beeping and no dye is coming out The current circuit is broken Check entire current path; if platinum is coated, remove coat with a razor blade; if the platinum wire has kinks in it, replace it; for air objective setups, ensure that there is enough liquid to create a current path between micropipette and ground
26–30 Dendrites appear beaded Too long time elapses between opening of chest cavity and start of perfusion See troubleshooting for step 26
A non-fixative (e.g. PB) was used before perfusate DO NOT use any non-fixative such as saline to “clear” the blood-vessels; this increases the duration of hypoxia of the brain and starts the apoptotic/necrotic process
47 Can only visualize and image dendrites that are less than 30 μm below the surface Coverslip is slanted Remount
Objective spring is broken Send in objective for repair
49 Dendrite appears to be moving across the screen during confocal imaging Nail polish has not had enough time to harden or not enough nail polish has been applied Because the recommended mounting media is aqueous and therefore never hardens, the stability of the slide prep relies on sufficient and hardened nail polish; wait 2–3 days before imaging; if movement is still seen, apply another coat of nail polish and wait another couple of days
49,58 Some dendrites or portions of dendrites appear fuzzy in the confocal image Air bubbles in the immersion oil Remove oil from slide and add a fresh amount of oil
Air bubbles in mounting media Remount
Dendrite is under a blood vessel Move on to the next segment; do not use for analysis
56–61 Problems with NeuronStudio Consult the online manual http://research.mssm.edu/cnic/help/ns/index.html