Table 1.

Overview of the used magnetic resonance sequences constituting the test battery for quality assurance testing

Sequence no. Name TR, ms TE, ms FA Bandwidth, Hz/Px Voxel Size, mm3 No. Slices Slice Orientation Scan Time, min:s Remarks
1 RF field 300 17 720° 130 1.2×1.2×5 1 Sagittal* 1:19 Spin echo
2 Signal-to-noise ratio 1,000 3.6 15° 260 1×1×5 12 Sagittal* 4:16 Gradient echo
2.5-mm Gap between slices
3 RF noise spikes 24 1 NA 130 2×2×10 1 Sagittal* 0:06
4 RF noise 1,400 1 NA 39 0.4×0.4×1 50 Transverse 5:58
5 Weisskoff (1996) EPI 2,000 30 90° 2,112 3.5×3.5×3.5 25 Transverse 10:44 Temporal stability estimation
6 Field map 704 5.11 7.57 60° 260 3.5×3.5×3.5 21 Transverse 1:33 B0 inhomogeneity estimation Magnitude and phase images
7 EPI with reference amplitude set to “0 V” 2,250 30 78° 2,112 3.5×3.5×3.5 21 Transverse 1:53 50% Phase oversampling 750-ms Gap between consecutive volumes
8 “Paradigm” EPI§ 2,250 30 78° 2,112 3.5×3.5×3.5 21 Transverse 1:53 50% Phase oversampling 750-ms Gap
TMS pulses were applied in TMS+EEG setupa Reference amplitude: 104.4 V
9 MPRAGE 2,250 2.6 200 1×1×1 192 Sagittal 8:26 T1-weighted anatomic image
*

Sagittal slice in line with main horizontal axis at center of transcranial magnetic stimulation (TMS) coil.

Radio frequency (RF) noise check in which the receiver chain of the system was set to a high receiver gain and symmetrically stepped through a bandwidth up to 250 kHz relative to the Larmor frequency of the system (i.e., 127.7 MHz) in steps of 10 kHz (i.e., 50 steps in total covering a range of 500 kHz).

50% Phase oversampling to reduce ghost artifacts of TMS coil (Bestmann et al. 2008a). §This echo-planar imaging (EPI) is similar to that used in the resting-state human data collection except that in the latter more volumes were collected and intensity was set to 80% of the stimulator output.

a

In TMS-EEG setup, each 9–27 s (average of 18 s), a TMS pulse (at 100% maximum stimulator output) was delivered in-between acquisitions. TR, repetition time; TE, echo time; FA, flip angle; NA, not applicable; B0, static magnetic field.