Table 3.

Examples of Device-Related Adverse Events in Hospitalized Children 1–12 Years of Age

Age Gender Patient Problem Reported Problem Pump Evaluation
35 months f Hyperglycemia The ACT button did not respond properly Pump operated OK but moisture on keypad

7 m Hypoglycemia, seizures Parent stated that she gave too much correction before going to bed Pump not returned
Use error

8 m Hypoglycemia Family uncomfortable with pump and wants replacement Pump tested OK

8 m Hyperglycemia, DKA Pump not correct size for body Battery was reported to be out of limit; pump tested OK

8 m Hyperglycemia, DKA Parent did not understand pump performance Pump not returned

8 m Hyperglycemia, DKA Parent not familiar with pump operation Pump not returned
Education needed

9 f Hyperglycemia, DKA Nondelivery during the middle of the night Pump not returned

9 f Hyperglycemia, DKA Numerous nondelivery alarms Pump not returned

10 m Hyperglycemia Hospital and family did not want to troubleshoot Pump not returned

11 f Hyperglycemia Bent cannula, which came out during sleep; two hospitalizations Pump not returned

11 m Hypoglycemia Parent activated the ACT button with pump in prime mode Pump not returned
Use error