Table 3.
DIN EN ISO 15197:2003 | Current draft revision of ISO 15197 | |||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
BG concentration <75 mg/dl | BG concentration ≥75 mg/dl | BG concentration <100 mg/dl | BG concentration ≥100 mg/dl | |||||||||||||||
BG system | Reference method | Within accuracy limits (±15 mg/dl and ±20%) | ±15 mg/dl | ±10 mg/dl | ±5 mg/dl | ±20% | ±15% | ±10% | ±5% | Within accuracy limits (±15 mg/dl and ±15%) | ±15 mg/dl | ±10 mg/dl | ±5 mg/dl | ±15% | ±10% | ±5% | ||
n | % | % | % | % | % | % | % | % | n | % | % | % | % | % | % | % | ||
Accu-Chek Active | HK | (200/200) | 100.0 | 100 | 100 | 61 | 100 | 100 | 100 | 81 | (200/200) | 100.0 | 100 | 100 | 72 | 100 | 100 | 79 |
Accu-Chek Aviva | HK | (200/200) | 100.0 | 100 | 100 | 87 | 100 | 99 | 91 | 64 | (198/200) | 99.0 | 100 | 97 | 80 | 99 | 91 | 64 |
Accu-Chek Aviva Nano | HK | (200/200) | 100.0 | 100 | 100 | 84 | 100 | 99 | 94 | 65 | (199/200) | 99.5 | 100 | 96 | 80 | 99 | 95 | 65 |
Accu-Chek Compact Plus | HK | (200/200) | 100.0 | 100 | 88 | 23 | 100 | 100 | 91 | 63 | (200/200) | 100.0 | 100 | 86 | 27 | 100 | 94 | 69 |
Accu-Chek Go | HK | (200/200) | 100.0 | 100 | 100 | 97 | 100 | 100 | 97 | 78 | (200/200) | 100.0 | 100 | 100 | 94 | 100 | 96 | 79 |
Accu-Chek Mobileb | HK | (199/200) | 99.5 | 98 | 98 | 75 | 100 | 100 | 97 | 67 | (199/200) | 99.5 | 98 | 98 | 73 | 100 | 96 | 66 |
Accu-Chek Mobileb | HK | (200/200) | 100.0 | 100 | 100 | 78 | 100 | 100 | 93 | 64 | (200/200) | 100.0 | 100 | 98 | 71 | 100 | 94 | 66 |
Accu-Chek Performab | HK | (199/200) | 99.5 | 100 | 100 | 79 | 99 | 99 | 93 | 67 | (199/200) | 99.5 | 100 | 95 | 72 | 99 | 94 | 68 |
Accu-Chek Performab | HK | (198/200) | 99.0 | 100 | 98 | 75 | 99 | 98 | 93 | 68 | (196/200) | 98.0 | 98 | 97 | 78 | 98 | 92 | 66 |
Accu-Chek Performa Nano | HK | (200/200) | 100.0 | 100 | 100 | 93 | 100 | 100 | 96 | 66 | (200/200) | 100.0 | 100 | 98 | 87 | 100 | 96 | 64 |
Bayer Contour usb | GOx | (194/200) | 97.0 | 100 | 84 | 61 | 96 | 88 | 63 | 31 | (182/200) | 91.0 | 90 | 69 | 45 | 91 | 68 | 34 |
Beurer GL32 | GOx | (192/200) | 96.0 | 80 | 48 | 30 | 100 | 98 | 90 | 56 | (189/200) | 94.5 | 85 | 62 | 40 | 99 | 91 | 56 |
Beurer GL40 | GOx | (198/200) | 99.0 | 97 | 95 | 53 | 99 | 94 | 78 | 43 | (192/200) | 96.0 | 98 | 90 | 52 | 95 | 78 | 42 |
Biocheck TD-4225 | GOx | (187/200) | 93.5 | 73 | 40 | 13 | 99 | 96 | 81 | 48 | (183/200) | 91.5 | 76 | 52 | 22 | 97 | 80 | 49 |
FreeStyle Freedom Lite | GOx | (200/200) | 100.0 | 100 | 100 | 98 | 100 | 100 | 98 | 91 | (200/200) | 100.0 | 100 | 100 | 98 | 100 | 98 | 90 |
FreeStyle Lite | GOx | (200/200) | 100.0 | 100 | 100 | 95 | 100 | 100 | 100 | 86 | (200/200) | 100.0 | 100 | 100 | 93 | 100 | 100 | 86 |
Futura Monometer | GOx | (182/200) | 91.0 | 93 | 75 | 23 | 91 | 78 | 59 | 28 | (165/200) | 82.5 | 90 | 68 | 27 | 79 | 59 | 26 |
GlucoCheck Classic | GOx | (191/200) | 95.5 | 100 | 93 | 60 | 94 | 86 | 66 | 38 | (177/200) | 88.5 | 97 | 84 | 55 | 85 | 66 | 38 |
GlucoCheck XL | GOx | (191/200) | 95.5 | 98 | 95 | 63 | 95 | 89 | 67 | 40 | (182/200) | 91.0 | 97 | 92 | 58 | 88 | 65 | 40 |
Glucohexal IIc | GOx | (162/200) | 81.0 | 61 | 24 | 8 | 86 | 74 | 57 | 28 | (143/200) | 71.5 | 50 | 21 | 7 | 80 | 64 | 32 |
GlucoRx (TD-4230) | GOx | (170/200) | 85.0 | 98 | 45 | 13 | 82 | 63 | 38 | 14 | (141/200) | 70.5 | 83 | 41 | 14 | 65 | 40 | 14 |
GlucoSmart Swing | GOx | (193/200) | 96.5 | 95 | 63 | 24 | 97 | 88 | 60 | 28 | (182/200) | 91.0 | 84 | 53 | 15 | 94 | 67 | 33 |
GlucoTel | GOx | (190/200) | 95.0 | 87 | 71 | 37 | 97 | 93 | 70 | 40 | (183/200) | 91.5 | 89 | 70 | 33 | 92 | 72 | 41 |
Gluco-test Plus+ TD-4230 | GOx | (198/200) | 99.0 | 100 | 98 | 75 | 99 | 94 | 80 | 45 | (190/200) | 95.0 | 96 | 94 | 76 | 95 | 79 | 42 |
iDia | GOx | (191/200) | 95.5 | 100 | 85 | 50 | 94 | 90 | 70 | 41 | (184/200) | 92.0 | 96 | 80 | 47 | 90 | 71 | 40 |
IME-DC Fidelity | GOx | (183/200) | 91.5 | 80 | 48 | 18 | 94 | 88 | 78 | 44 | (175/200) | 87.5 | 72 | 43 | 18 | 94 | 85 | 48 |
iXell | GOxd | (199/200) | 99.5 | 100 | 100 | 70 | 99 | 91 | 77 | 42 | (185/200) | 92.5 | 100 | 98 | 68 | 89 | 75 | 38 |
iXell OLED | GOxd | (198/200) | 99.0 | 95 | 78 | 45 | 100 | 98 | 81 | 42 | (194/200) | 97.0 | 97 | 85 | 52 | 97 | 79 | 39 |
microdot+ | GOx | (198/200) | 99.0 | 97 | 97 | 71 | 99 | 94 | 81 | 43 | (190/200) | 95.0 | 98 | 95 | 62 | 94 | 83 | 43 |
OneTouch Verio | GOx | (199/200) | 99.5 | 100 | 87 | 34 | 99 | 96 | 90 | 62 | (198/200) | 99.0 | 97 | 78 | 40 | 100 | 95 | 64 |
OneTouch Verio Pro | GOx | (193/200) | 96.5 | 93 | 63 | 20 | 98 | 90 | 70 | 38 | (183/200) | 91.5 | 88 | 53 | 21 | 93 | 75 | 40 |
Pura | HK | (200/200) | 100.0 | 100 | 92 | 55 | 100 | 100 | 75 | 30 | (200/200) | 100.0 | 100 | 95 | 48 | 100 | 74 | 30 |
SeniorLine GM210 | GOx | (144/200) | 72.0 | 10 | 0 | 0 | 88 | 72 | 48 | 26 | (120/200) | 60.0 | 15 | 0 | 0 | 79 | 55 | 29 |
Wellion CALLA Light | GOx | (182/200) | 91.0 | 68 | 33 | 10 | 97 | 85 | 68 | 48 | (165/200) | 82.5 | 68 | 33 | 13 | 89 | 73 | 52 |
200 results from 100 subjects were evaluated.
Accu-Chek Mobile and Accu-Chek Performa were both tested with different test strip chemistries. The test strip chemistry was either maltose dependent (test strips evaluated first) or maltose independent (test strips evaluated second).
GlucoHexal test strip lot was recalled from the market in June 2010, after at least 11 months availability on the market.
Information was not available.