Table 3.

Relative Bias and Limits of Agreement according to Bland and Altmana

BG system Reference method Average relative bias (%) Test strip lot Relative bias (%) Lower limit of agreement (%) Upper limit of agreement (%)
Accu-Chek Aviva (system A) HK -0.4 A1 -0.7 -11.8 10.5

A2 1.1 -8.3 10.5

A3 -1.0 -10.5 8.5

A4 -0.9 -10.0 8.3

FreeStyle Lite (system B) GOx -4.0 B1 -0.8 -7.6 5.9

B2 -6.0 -22.0 10.1

B3 -1.4 -11.8 9.0

B4 -7.7 -20.2 4.8

GlucoCheck XL (system C) GOx 2.8 C1 4.6 -13.3 22.4

C2 2.7 -16.9 22.4

C3 2.3 -20.8 25.4

C4 1.5 -17.7 20.6

Pura/mylife Pura (system D) HK -9.0 D1 -6.7 -19.5 6.1

D2 -17.3 -33.9 -0.7

D3 -4.3 -16.4 7.8

D4 -7.6 -19.2 4.0

OneTouch Verio Pro (system E) GOx 8.8 E1 8.5 -4.2 21.2

E2 9.5 -4.0 22.9

E3 8.7 -3.0 20.4

E4 8.5 -5.0 22.0
a

Reference methods (GOx or HK) as mentioned in the manufacturer's labeling. For the calculation of the relative bias of each test strip lot, data of 200 blood samples (BG concentrations <50 to >400 mg/dl) were included. The average relative bias of a BG system is the mean value of the four corresponding test strip lots' relative biases.