Table 1.

Clinical goals for organs at risk to be satisfied by H&N plans. In difficult cases, the physician might allow the relaxation of a goal to the number observed in parenthesis. The achievable parotid gland mean dose constraint is estimated per patient using the heuristic proposed by Hunt et al.19 In addition to the structures in the table, hot spots more than ≈1 cm outside of the envelope of the lowest dose PTV are not allowed.

Structure Constraint
Cord Dmax ⩽ 45 Gy (50 Gy)
Brainstem Dmax ⩽ 50 Gy (60 Gy)
Chiasm Dmax ⩽ 54 Gy (60 Gy)
Mandible excluding all PTVs Dmax ⩽ 104% of low PTV Rx
Cochlea Dmax ⩽ 50 Gy
Brachial plexus Dmax ⩽ 65 Gy
Lens Dmax ⩽ 5 Gy (10 Gy)
Eyes/retina Dmax ⩽ 45 Gy
Optic nerve Dmax ⩽ 54 Gy
Larynx Dmean ⩽ 40 − 45 Gy
Parotid Dmean ⩽ 26 Gy
Oral cavity Dmean ⩽ 35 − 40 Gy