Table 6.
PLR rates for the on-body measurements with IPANEMA generation 2.5 nodes. Please note that the ambient hygro-/thermometer was not used during the outdoor measurements. The number of indoor transmitted packets is larger because more indoor measurements were performed than climate chamber and outdoor measurements.
| Sensor |
Indoor PLR [%] |
Outdoor PLR [%] |
Climate PLR [%] |
|
|---|---|---|---|---|
| ECG | Mean | 2.10 | 3.61 | 0.52 |
| Std. | 7.48 | 7.25 | 1.66 | |
| n | 2,155,776 | 396,735 | 1,677,680 | |
| ACC1FOOT | Mean | 3.61 | 4.18 | 2.75 |
| Std. | 6.63 | 6.06 | 3.98 | |
| n | 471,164 | 82,589 | 346,555 | |
| ACC2BACK | Mean | 2.04 | 2.07 | 1.10 |
| Std. | 5.49 | 4.42 | 1.65 | |
| n | 469,639 | 67,776 | 345,739 | |
| HYGRO1SKIN | Mean | 2.71 | 3.71 | 1.68 |
| Std. | 5.84 | 8.67 | 1.90 | |
| n | 19,207 | 1366 | 15,617 | |
| HYGRO2ROOM | Mean | 2.11 | - | 1.45 |
| Std. | 4.92 | - | 1.71 | |
| n | 21,977 | - | 16,099 |