Published in final edited form as: Ultrasound Med Biol. 2012 Aug 21;38(10):1784–1798. doi: 10.1016/j.ultrasmedbio.2012.06.013

Table 6.

Comparing errors between true pressures obtained with Millar catheter and the pressures obtained with SHAPE data using cross-validation approach

Incident Acoustic Power (%) Using data from MIP image (SHMIP) SHAPE vs. catheter pressures
Using data from from mean signal of all frames (SHAll_Frames) SHAPE vs. catheter pressures
Mean Difference (mmHg) 95 % Confidence Interval of Difference (mmHg) p-value Mean Difference (mmHg) 95 % Confidence Interval of Difference (mmHg) p-value
a) For all data
20 -0.03 -3.0 to 3.0 0.982 -0.15 -3.4 to 3.1 0.924
40 0.08 -2.7 to 2.9 0.951 -0.09 -3.3 to 3.1 0.952

b) For baseline data only (before PH)
20 3.47 -0.1 to 7.0 0.053 3.50 -0.9 to 7.9 0.110
40 3.23 -0.3 to 6.8 0.070 3.64 -0.9 to 8.2 0.103

c) For PH data only
20 -3.5 -8.0 to 0.9 0.109 -3.8 -8.0 to 0.4 0.072
40 -3.1 -7.0 to 0.9 0.113 -3.8 -7.9 to 0.2 0.062