Table 1.
Isotope |
VOI |
Quantitative error |
Changes induced by map inaccuracies (rotation, shift and/orμchange) with respect to references |
|||||||
---|---|---|---|---|---|---|---|---|---|---|
0° |
0° |
0° |
0° |
15° |
0° |
0° |
15° |
|||
1 mmX |
3 mmX |
1 mmY |
3 mmY |
0 mm |
0 mm |
0 mm |
3 mmX |
|||
1.0 μ | 1.0 μ | 1.0 μ | 1.0 μ | 1.0 μ | 0.9 μ | 1.1 μ | 1.1 μ | |||
125I |
1 |
−20.4 |
0.1 |
0.4 |
−0.5 |
−4.0 |
0.0 |
−4.9 |
5.2 |
5.3 |
2 |
−11.8 |
0.7 |
1.1 |
−0.2 |
−1.0 |
−0.2 |
−4.8 |
5.0 |
6.2 |
|
3 |
−7.9 |
−1.1 |
−4.0 |
−1.5 |
−4.5 |
−1.4 |
−4.2 |
4.3 |
−0.8 |
|
201Tl |
1 |
−5.8 |
0.0 |
0.1 |
−0.1 |
−1.5 |
0.0 |
−2.6 |
2.7 |
2.7 |
2 |
6.1 |
−0.4 |
−1.4 |
0.0 |
−0.3 |
−0.1 |
−2.5 |
2.6 |
3.0 |
|
3 |
4.4 |
0.2 |
−0.3 |
0.6 |
1.4 |
−0.7 |
−2.3 |
2.3 |
0.0 |
|
99mTc |
1 |
−2.0 |
0.1 |
0.2 |
−0.2 |
−1.7 |
0.0 |
−2.2 |
2.2 |
2.1 |
2 |
6.1 |
−0.3 |
−1.1 |
−0.1 |
−0.3 |
−0.1 |
−2.1 |
2.2 |
2.4 |
|
3 |
−8.3 |
0.4 |
−0.2 |
0.5 |
1.3 |
−0.2 |
−1.8 |
1.9 |
0.2 |
|
111In | 1 |
−4.7 |
0.0 |
0.1 |
−0.2 |
−1.4 |
0.0 |
−2.1 |
2.1 |
2.2 |
2 |
−3.5 |
−0.4 |
−1.2 |
0.0 |
−0.3 |
−0.1 |
−2.0 |
2.0 |
2.5 |
|
3 | 3.2 | 0.0 | −1.1 | 0.4 | 0.9 | −0.9 | −1.9 | 1.9 | 0.6 |
Quantitative errors (in %) of VOI activities in reference phantom SPECT images were calculated with respect to dose calibrator measurements (1st data column), and relative changes (in %) in VOI activities due to attenuation map inaccuracies were calculated with respect to reference images (2nd to 9th data column).