Skip to main content
. 2013 Jan 31;3:7. doi: 10.1186/2191-219X-3-7

Table 1.

Quantitative errors and relative changes of VOI activities

Isotope
VOI
Quantitative error
Changes induced by map inaccuracies (rotation, shift and/orμchange) with respect to references




15°


15°
1 mmX
3 mmX
1 mmY
3 mmY
0 mm
0 mm
0 mm
3 mmX
      1.0μ 1.0μ 1.0μ 1.0μ 1.0μ 0.9μ 1.1μ 1.1μ
125I
1
−20.4
0.1
0.4
−0.5
−4.0
0.0
−4.9
5.2
5.3
2
−11.8
0.7
1.1
−0.2
−1.0
−0.2
−4.8
5.0
6.2
3
−7.9
−1.1
−4.0
−1.5
−4.5
−1.4
−4.2
4.3
−0.8
201Tl
1
−5.8
0.0
0.1
−0.1
−1.5
0.0
−2.6
2.7
2.7
2
6.1
−0.4
−1.4
0.0
−0.3
−0.1
−2.5
2.6
3.0
3
4.4
0.2
−0.3
0.6
1.4
−0.7
−2.3
2.3
0.0
99mTc
1
−2.0
0.1
0.2
−0.2
−1.7
0.0
−2.2
2.2
2.1
2
6.1
−0.3
−1.1
−0.1
−0.3
−0.1
−2.1
2.2
2.4
3
−8.3
0.4
−0.2
0.5
1.3
−0.2
−1.8
1.9
0.2
111In 1
−4.7
0.0
0.1
−0.2
−1.4
0.0
−2.1
2.1
2.2
2
−3.5
−0.4
−1.2
0.0
−0.3
−0.1
−2.0
2.0
2.5
3 3.2 0.0 −1.1 0.4 0.9 −0.9 −1.9 1.9 0.6

Quantitative errors (in %) of VOI activities in reference phantom SPECT images were calculated with respect to dose calibrator measurements (1st data column), and relative changes (in %) in VOI activities due to attenuation map inaccuracies were calculated with respect to reference images (2nd to 9th data column).