Skip to main content
. 2012 May 30;16(3):R98. doi: 10.1186/cc11366

Table 4.

Agreement between trends in CO, GEDV and EVLW measurements by the VolumeView/EV1000™ and the PiCCO2™ system.

four quadrant plot polar plots


conctot conc10% concordance radial LOA
CO 87.6% 98.5% 100% 0.31 ml/minute
GEDV 81.5% 95.1% 99.7% 81 ml
EVLW 88.6% 97.7% 99.7% 40 ml

CO, cardiac output; GEDV, global enddiastolic volume; EVLW, extravascular lung water; conctot, concordance rate including all data points; conc10%, concordance rate using a 10% exclusion zone; LOA, limits of agreement.