Published in final edited form as: J Magn Reson Imaging. 2012 Nov 21;37(6):1419–1426. doi: 10.1002/jmri.23936

Table II.

Statistical Comparison of Various Flows by 4-D phase contrast (n=22). Lower limit of agreement represents the mean difference −1.96 standard deviations and upper limit of agreement represents the mean difference +1.96 standard deviations

Upper Body Lower Body Whole Body Whole Body Whole Body Lower Body
Flow A AA- DA IVC SVC + IVC SVC + IVC AA DAD
Flow B SVC DA AA PT PT DA
Pearson Correlation Coefficient 0.89 0.96 0.97 0.94 0.97 0.98
Paired T-test p values 0.06 0.09 0.01 0.38 0.89 0.72
Bland-Altman Absolute
Mean Difference (L/min) −0.1 0.1 0.2 0.1 −0.1 0
Lower Limit of Agreement (L/min) −0.5 −0.3 −0.3 −0.6 −0.6 −0.2
Upper Limit of Agreement (L/min) 0.3 0.4 0.6 0.7 0.4 0.3
Bland-Altman Relative
Mean Difference −6% 6% 7% 2% −4% 1%
Lower Limit of Agreement −44% −22% −13% −21% −22% −18%
Upper Limit of Agreement 32% 34% 26% 26% 14% 21%

AA= Ascending Aorta, DA= Descending Aorta after last arch vessel, DAD= Descending Aorta just above diaphragm, PT= Pulmonary Trunk, SVC= Superior Vena Cava, IVC= Inferior Vena Cava