Skip to main content
. 2013 Feb 26;84(1):98–105. doi: 10.3109/17453674.2013.765625

Table 3.

Absolute translations (in mm) of the clinical RSA examinations at 2-year follow-up. There was no statistically significant difference between the groups (p > 0.2)

Axis: DLC all-poly cup
(n = 10 a)
Electra screw cup
(n = 10 a)
x y z TT b x y z TT b
Mean (mm) 0.09 0.12 0.14 0.24 0.06 0.12 0.14 0.19
SD (mm) 0.06 0.05 0.14 0.10 0.07 0.14 0.10 0.16
Minimum 0.002 0.04 0.03 0.11 0.02 0.001 0.001 0.05
Maximum 0.17 0.22 0.42 0.44 0.25 0.49 0.26 0.61

a Total translation (TT) values are given without the 2 outliers at 2 years (1 loose/revised cup and 1 suspected loose cup; both with TT of > 1 mm).

b The total translation was calculated using the 3D Pythagorean theorem.