Table 2.

Comparison of quantification efficiencies between LTQ and LTQ-Orbitrap for 11 different 6-step MudPIT experiments.

Sample # Spectra Identified LTQ # Spectra Quantified * LTQ Quantification Efficiency Orbitrap # Spectra Quantified* Orbitrap Quantification Efficiency
1:1 528 266 0.50 361 0.68
1:1 1267 595 0.47 741 0.58
1:1 1637 832 0.51 1336 0.82
5:1 3099 328 0.11 1662 0.54
5:1 1121 127 0.11 591 0.53
5:1 2885 464 0.16 1475 0.51
10:1 4019 378 0.09 1856 0.46
10:1 3900 350 0.09 1683 0.43
10:1 2674 180 0.07 839 0.31
50:1 2283 1 0.00 210 0.09
100:1 2767 4 0.00 224 0.08
*

criteria used for quantification include an r2 value > 0.5 and accuracy within 2 fold of the expected ratio. In addition, statistical outliers were removed using grubbs test with an outlier threshold p-value of 0.1.