Table 1.

Average values and standard deviation for central corneal thickness and anterior chamber depth measurements obtained with the ultrasound devices and Sirius® for three runs of measurements

n = 50 Central corneal thickness Anterior chamber depth



Measure Sirius (mean ± SD) (μm) SP100 (mean ± SD) (μm) Sirius (mean ± SD) (mm) US800 (mean ± SD) (mm)
1 546.5 ± 39.43 540.6 ± 34.46 2.95 ± 0.33 3.37 ± 0.30
2 545.4 ± 38.46 541.4 ± 36.07 2.97 ± 0.31 3.37 ± 0.32
3 545.5 ± 38.63 542.0 ± 35.01 2.96 ± 0.30 3.34 ± 0.30
P 0.987* 0.980* 0.852** 0.964*

Notes:P, statistical significance for differences among the three test runs as obtained using

*

analysis of variance test;

**

Kruskal-Wallis test.

Abbreviation: SD, standard deviation.