. 2013 Feb 26;7:417–422. doi: 10.2147/OPTH.S35121

Table 3.

Mean difference, standard deviation, level of statistical significance, and limits of agreement between each set of measurement techniques to be compared at the 95% confidence interval

Mean SD P Limits of agreement

Mean difference −1.96*SD Mean difference +1.96*SD
CCT (μm)
 Sirius versus SP100 4.68 10.47 0.003 −15.84 25.20
ACD (mm)
 Sirius versus US800 −0.40 0.16 <0.001 −0.72 −0.07
 Sirius + CCT* versus US800 0.15 0.16 <0.001 −0.16 0.45
 Sirius versus US800 – CCT** 0.15 0.16 <0.001 −0.16 0.45

Notes:

*

Sirius + CCT represents the sum of ACD and CCT obtained with Sirius;

**

US800 – CCT represents the value of ACD obtained with US800 ultrasound biometry minus the CCT obtained with ultrasound pachymetry.

Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness; SD, standard deviation.