Table 3.
Mean | SD | P | Limits of agreement | ||
---|---|---|---|---|---|
|
|||||
Mean difference −1.96*SD | Mean difference +1.96*SD | ||||
CCT (μm) | |||||
Sirius versus SP100 | 4.68 | 10.47 | 0.003 | −15.84 | 25.20 |
ACD (mm) | |||||
Sirius versus US800 | −0.40 | 0.16 | <0.001 | −0.72 | −0.07 |
Sirius + CCT* versus US800 | 0.15 | 0.16 | <0.001 | −0.16 | 0.45 |
Sirius versus US800 – CCT** | 0.15 | 0.16 | <0.001 | −0.16 | 0.45 |
Notes:
Sirius + CCT represents the sum of ACD and CCT obtained with Sirius;
US800 – CCT represents the value of ACD obtained with US800 ultrasound biometry minus the CCT obtained with ultrasound pachymetry.
Abbreviations: ACD, anterior chamber depth; CCT, central corneal thickness; SD, standard deviation.