Skip to main content
. 2013 Mar 5;8(3):e57631. doi: 10.1371/journal.pone.0057631

Table 1. Inter-assay % Bias and precision (% CV) of lyso-Gb3 quality control mouse tissue samples.

QC (Actual Conc.) Kidney Skin Heart Plasma
Low1(2.00 ng/mL) Mean ± SD 2.16±0.306 2.02±0.233 2.12±0.122 2.15±0.210
% CV 14.2 11.5 5.77 9.78
% Bias 7.89 1.00 6.17 7.36
Low2(4.00 ng/mL) Mean ± SD 4.10±0.327 4.18±0.517 4.23±0.413 4.09±0.616
% CV 8.0 12.4 9.78 15.1
% Bias 2.50 4.43 5.67 2.25
Mid(80.0 ng/mL) Mean ± SD 84.9±8.43 82.3±5.51 92.5±3.91 79.8±12.6
% CV 9.93 6.70 4.23 15.8
% Bias 6.08 2.88 15.6 −0.250
High(160 ng/mL) Mean ± SD 162±19.7 167±13.9 181±27.2 151±11.8
% CV 12.2 8.34 15.0 7.83
% Bias 1.32 4.50 13.2 −5.70

Quality control samples were prepared in wild-type mouse plasma or tissue homogenates at four concentration levels and analyzed on three separate days. Six to 10 replicates at each concentration level were used for inter-assay determination. Conc., concentration; SD, Standard Deviation.