Table 2. Inter-assay % Bias and precision (% CV) of lyso-Gb3 quality control human plasma samples.
QC (actual conc.) | Low1 (2.00) | Low2 (4.00) | Mid1 (40.0) | Mid2 (80.0) | High (200) |
Mean ± SD (ng/mL) | 2.07±0.249 | 3.75±0.617 | 41.2±5.20 | 75.0±6.94 | 200±19.7 |
% CV | 12.0 | 16.4 | 12.6 | 9.25 | 9.84 |
% Bias | 3.60 | −6.16 | 2.95 | −6.19 | 0.15 |
Quality control samples were prepared in normal human plasma at five concentration levels and analyzed on four separate days. Twenty replicates at each concentration level were used for inter-assay determination. Conc., concentration (ng/mL); SD, Standard Deviation.