Skip to main content
. 2013 Jan 15;405(9):2943–2955. doi: 10.1007/s00216-012-6674-3

Table 5.

Summary of NUSIMEP-7 results

NUSIMEP-7 test sample 235U/238U CRM for external correction RU (k = 2) (%) RDb (%) z scorec zeta scorec Cluster size/data points
‘Single’ deposition 0.00898(90) IRMM-184 10.0 −1.0 −2.04 −0.21 144
0.2130(10) CRM U500 0.5 a a a 119
‘Double’ deposition 0.0090(11) IRMM-184 12.2 −0.7 −1.60 −0.13 29
0.0332(10) CRM U030-A 3.0 −2.6 −5.55 −1.89 633
0.2096(11) CRM U500 0.5 a a a 122

aContamination

bRelative difference to the certified value

cInterpretation of scores: |score| ≤ 2, satisfactory result; 2 < |score| < 3, questionable result; |score| > 3, unsatisfactory result