Table 2.

4D kinematic CT wrist scanning parameters compared to static high resolution 3D scan

Scan Mode Slice thickness Range kV mA Rotation Time Acquisition Time Acquisition Interval CTDIvol (mGy) DLP (mGy-cm) Effective Dose
Static high-resolution 3D scan
Volume 0.5 mm × 160 80 mm 120 100 0.5 s NA NA 4.1 32.4 0.67
4D Kinematic Dynamic Volume (Large patient or with metal in situ)
Dy-Volume 0.5 mm × 160 80 mm 100 100 0.5 s 5.5 s Continuous 26.0 207.7 0.17
4D Kinematic Dynamic Volume (Average patient)
Dy-Volume 0.5 mm × 160 80 mm 80 80 0.5 s 5.5 s Continuous 10.5 84.3 0.07

NA indicates not applicable; CTDIvol, CT dose index volume.