Skip to main content
. Author manuscript; available in PMC: 2013 Mar 10.
Published in final edited form as: AJR Am J Roentgenol. 2013 Jan;200(1):24–34. doi: 10.2214/AJR.12.9543

TABLE 1.

Typical Sequences Used and Minor Practical Requirements

Sequence T1-Weighted (DCE) T2*-Weighted (DSC) ASL

SPGR/MP RAGE/FLASH/FFE (Typically 3D) GRE Echo-Planar Imaging (2D Multislice) GRE Echo-Planar Imaging (2D Multislice)

Temporal resolution ~3–6 s ~1–2 s 3–5 s
Total acquisition time 3–5 min 2 min 3–5 min
Spatial resolution 1-mm in-plane × 5-mm slices 2-mm in-plane × 5-mm slices 3-mm in-plane × 5-mm slices
Geometric artifact Low impact Prone to problems at the skull base Prone to problems at the skull base
Model parameters ktrans, vp, ve, IAUC CBV, CBF, MTT CBF

Note—DCE = dynamic contrast-enhanced, DSC = dynamic susceptibility contrast-enhanced, ASL = arterial spin labeling, SPGR = spoiled gradient-recalled echo, MP RAGE = magnetization prepared rapid gradient echo, FFE = fast-field echo, GRE = gradient-recalled echo, ktrans = transfer constant, vp = fractional volume of the plasma space, ve = fractional volume of the extravascular-extracellular space, IAUC = initial area under the contrast agent concentration-time curve, CBV = cerebral blood volume, CBF = cerebral blood flow, MTT = mean transit time.