Table 5.

Measures of validity of the EasyOne spirometer compared to laboratory spirometry in the clinical setting with spirometry-naive participants.

Mean difference (95% confidence intervals) 95% limits of agreement Slope of line in Bland-Altman plots of absolute difference vs. mean value
β P-value
FVC* (L) −0.12 (−0.19, −0.04) −0.48, 0.25 −0.05 0.14
FEV1 (L) −0.17 (−0.23, −0.11) −0.45, 0.11 −0.06 0.04
FEV1/FVC −0.02 (−0.03, −0.01) −0.05, 0.02 0.06 0.26
FEV6 (L) −0.14 (−0.21, −0.07) −0.48, 0.19 −0.05 0.14
FEF25–75§ (L) −0.27 (−0.40, −0.13) −0.93, 0.40 −0.04 0.37
PEF (L/sec) 0.18 (−0.10, 0.45) −1.18, 1.54 0.05 0.45
*

FVC = forced vital capacity

FEV1 = forced expiratory volume in one second

FEV6 = forced expiratory volume in six seconds

§

FEF25–75 = forced expiratory flow between 25% and 75% of vital capacity

PEF = peak expiratory flow rate