Table 1.

Cone Angle*

3.5°
Image Mid-plane Spatial Resolution (FWHM - mm) 1.61 1.71 1.71
Resolution (mm) at ± 15 mm above, below image plane 1.8 (1.8) 2.2 (2.2) 2.6 (2.5)
Sensitivity cps/kBq– 18F 0.0730 (0.0740) 0.234 (0.227) 0.465 (0.468)
Uniformity Fraction (dimensionless) 0.75 (0.76) 0.58 (0.58) 0.43 (0.48)
Total FOV 176Lu Background Count Rate (cps) field- flood corrected 1.3 4.0 7.6
*

Energy window: 250–650 keV, all entries. Measured values are without parentheses, calculated values are with parentheses. Calculated values for Resolution and Uniformity Fraction where obtained from Equations 1 and 3, respectively. Calculated values for Sensitivity were obtained from Equation 2 with α= 0.64 and ε= 0.77.